Do not pump:
-
unstable substances
- substances which react explosively under
impact
(mechanical stress) without air
- substances which react explosively when being ex-
posed to
elevated temperatures without air,
- substances subject to
auto-ignition,
- substances which are
inflammable without air
-
explosive substances.
➨
The pumps are
not approved for operation below
ground.
Do not operate the pump below ground.
•
Do not pump substances which may form depos-
its inside the pump. The pumps are not suitable for
pumping substances which may form deposits inside
the pump. Deposits and condensate in the pump may
lead to increased temperatures even to the point of
exceeding the maximum permitted temperatures.
• Check the inlet and outlet of the pump, if there is a
danger of forming
deposits inside the pump, e.g., in
the pump chambers (the pump chamber is the part
between diaphragm and head cover. See section “Re-
placing diaphragms and valves”, pg. 49). Inspect the
pump chambers regularly and clean if necessary.
•
Consider interactions and chemical reactions of
the pumped media. Ensure that the materials of the
pump’s wetted parts are compatible with the pumped
substances, see section “Technical data”, pg. 24.
When changing the substances pumped, we recom-
mend purging the pump with air or inert gas prior to
changing the pumped media. Purging the pump will
pump out residues and it will reduce the possibility of
reactions of the pumped substances with each other
and with the pump’s materials.
- MV 10C VARIO-B without controller (o.C.) ME 4C NT VARIO without controller (o.C.) MD 4C NT VARIO without controller (o.C.) MD 12C VARIO-B without controller (o.C.) PC 3012 VARIO without controller (o.C.) PC 3010 VARIO without controller (o.C.) PC 3004 VARIO without controller (o.C.) PC 3003 VARIO without controller (o.C.) PC 3002 VARIO without controller (o.C.)