beautypg.com

2 compute r0 & alpha 38, 3 compute beta (cold side only) 38, Compute r0 & alpha – ThermoWorks 9009-B INDUSTRIAL DUAL-BLOCK CALIBRATOR User Manual

Page 41: Compute beta (cold side only), 2 compute r0 & alpha, 3 compute beta (cold side only)

background image

C

T

T

T

T

= ⎡

⎣⎢


⎦⎥


⎣⎢


⎦⎥

− ⎡

⎣⎢


⎦⎥


⎣⎢


4

4

3

3

100

1

100

100

1

100 ⎥

D

T

T

T

T

= ⎡

⎣⎢


⎦⎥


⎣⎢


⎦⎥

− ⎡

⎣⎢


⎦⎥


⎣⎢


3

3

2

2

100

1

100

100

1

100 ⎥

E

R

R

=

4

3

F

R

R

=

3

2

delta

AF

BE

DE

CF

=

Where:

T

1-4

- Measured temperature using thermometer.

R

1-4

- Value of R from display of 9009 (Press SET and DOWN at the same

time.)

where

T

1

and R

1

are the measured temperature and resistance at –15 °C

T

2

and R

2

are the measured temperature and resistance at 0 °C

T

3

and R

3

are the measured temperature and resistance at 60 °C

T

4

and R

4

are the measured temperature and resistance at 110°C

10.2.2

Compute R0 & ALPHA

a

T

delta

T

T

1

2

2

2

100

1

100

=

+


⎣⎢


⎦⎥


⎣⎢


⎦⎥

a

T

delta

T

T

3

4

4

4

100

1

100

=

+


⎣⎢


⎦⎥


⎣⎢


⎦⎥

rzero

R a

R a

a

a

=

4

1

2

3

1

3

alpha

R

R

R a

R a

=

2

4

4

1

2

3

Where:

delta is the new value of DELTA computed above

10.2.3

Compute BETA (Cold Side Only)

x

T

= ⎡

⎣⎢


⎦⎥

1

100

1

9009 Industrial Dual-Block Calibrator
User’s Guide

38