2 compute r0 and alpha 40, Compute r0 and alpha, 2 compute r0 and alpha – ThermoWorks 9103 HART SCIENTIFIC FIELD DRY-WELL TEMPERATURE CALIBRATORS User Manual

Page 42

background image

B

T

T

=

3

2

C

T

T

T

T

= ⎡

⎣⎢


⎦⎥


⎣⎢


⎦⎥

− ⎡

⎣⎢


⎦⎥


⎣⎢


4

4

3

3

100

1

100

100

1

100 ⎥

D

T

T

T

T

= ⎡

⎣⎢


⎦⎥


⎣⎢


⎦⎥

− ⎡

⎣⎢


⎦⎥


⎣⎢


3

3

2

2

100

1

100

100

1

100 ⎥

E

R

R

=

4

3

F

R

R

=

3

2

delta

AF

BE

DE

CF

=

T

1-4

- Measured temperature using thermometer.

R

1-4

- Value of R from display of 9103 (Press SET and DOWN at the same

time.)

where

T

1

and R

1

are the measured temperature and resistance at –25°C

T

2

and R

2

are the measured temperature and resistance at 0°C

T

3

and R

3

are the measured temperature and resistance at 60°C

T

4

and R

4

are the measured temperature and resistance at 125°C

10.2.2

Compute R0 and ALPHA

a

T

delta

T

T

1

2

2

2

100

1

100

=

+


⎣⎢


⎦⎥


⎣⎢


⎦⎥

a

T

delta

T

T

3

4

4

4

100

1

100

=

+


⎣⎢


⎦⎥


⎣⎢


⎦⎥

rzero

R a

R a

a

a

=

4

1

2

3

1

3

alpha

R

R

R a

R a

=

2

4

4

1

2

3

delta is the new value of DELTA computed above.

9103 Dry-Well
User’s Guide

40