3 scan 22, 1 scan control 22, 2 scan rate 22 – ThermoWorks 9103 HART SCIENTIFIC FIELD DRY-WELL TEMPERATURE CALIBRATORS User Manual
Page 26: 4 temperature display hold 22, Scan, Scan control, Scan rate, Temperature display hold, 3 scan, 4 temperature display hold

7.3
Scan
The scan rate can be set and enabled so that when the set-point is changed the
calibrator heats or cools at a specified rate (degrees per minute) until it reaches
the new set-point. With the scan disabled the calibrator heats or cools at the
maximum possible rate.
7.3.1
Scan Control
The scan is controlled with the scan on/off function that appears in the main
menu after the set-point function.
Sc=OFF
Scan function off
Press “UP” or “DOWN” to toggle the scan on or off.
Sc=On
Scan function on
Press “SET” to accept the present setting and continue.
S
Accept scan setting
7.3.2
Scan Rate
The next function in the main menu is the scan rate. The scan rate can be set
from .1 to 99.9°C/minute. The maximum scan rate however is actually limited
by the natural heating or cooling rate of the instrument. This rate is often less
than 100°C/minute, especially when cooling.
The scan rate function appears in the main menu after the scan control function.
The scan rate units are in degrees per minute, degrees C or F depending on the
selected units.
Sr=10.0
Scan rate in °C/min.
Press “UP” or “DOWN” to change the scan rate.
Sr= 2.0
New scan rate
Press “SET” to accept the new scan rate and continue.
S
Accept scan rate
7.4
Temperature Display Hold
The calibrator has a display hold function which allows action of an external
switch to freeze the displayed temperature and stop the set-point from scan-
ning. This function is useful for testing thermal switches and cutouts. This sec-
9103 Dry-Well
User’s Guide
22