Scan

The scan rate can be set and enabled so that when the set-point is changed the
calibrator heats or cools at a specified rate (degrees per minute) until it reaches
the new set-point. With the scan disabled the calibrator heats or cools at the
maximum possible rate.

7.3.1

Scan Control

The scan is controlled with the scan on/off function that appears in the main
menu after the set-point function.

Sc=OFF

Scan function off

Press “UP” or “DOWN” to toggle the scan on or off.

Sc=On

Scan function on

Press “SET” to accept the present setting and continue.

S

Accept scan setting

7.3.2

Scan Rate

The next function in the main menu is the scan rate. The scan rate can be set
from .1 to 99.9°C/minute. The maximum scan rate however is actually limited
by the natural heating or cooling rate of the instrument. This rate is often less
than 100°C/minute, especially when cooling.

The scan rate function appears in the main menu after the scan control function.
The scan rate units are in degrees per minute, degrees C or F depending on the
selected units.

Sr=10.0

Scan rate in °C/min.

Press “UP” or “DOWN” to change the scan rate.

Sr= 2.0

New scan rate

Press “SET” to accept the new scan rate and continue.

S

Accept scan rate

7.4

Temperature Display Hold

The calibrator has a display hold function which allows action of an external
switch to freeze the displayed temperature and stop the set-point from scan-
ning. This function is useful for testing thermal switches and cutouts. This sec-