Table 2 – ThermoWorks 9103 HART SCIENTIFIC FIELD DRY-WELL TEMPERATURE CALIBRATORS User Manual

Page 39

background image

35

8 Digital Communication Interface

Interface Commands

Command Description

Command
Format

Command
Example

Returned

Returned
Example

Acceptable
Values

Read R0 calibration parameter

r[0]

r

r0: 999.999

r0: 100.578

Set R0 calibration parameter to

n r[0]=n

r=100.324

90 to 110

Read ALPHA calibration
parameter

al[pha]

al

al: 9.9999999

al: 0.0038573

Set ALPHA calibration parameter
to

n

al[pha]=n

al=0.0038433

.002 to .005

Read DELTA calibration
parameter

de[lta]

de

de:9.99999

de: 1.507

Set DELTA calibration parameter de[lta]=n

de=1.3742

0–3.0

Read BETA calibration parameter be[ta]

be

be:9.999

be:0.342

Set BETA calibration parameter

be[ta]=n

be=0.342

–100.0 to 100.0

Functions not on menu

Read firmware version number

*ver[sion]

*ver

ver.9999,9.99

ver.9103,2.00

Read structure of all commands

h[elp]

h

list of commands

Read all operating parameters

all

all

list of parameters

Legend:

[] Optional Command data

{} Returns either information

n Numeric data supplied by user

9 Numeric data returned to user

x Character data returned to user

Note:

When DUPLEX is set to FULL and a command is sent to READ, the command is returned followed by a
carriage return and linefeed. Then the value is returned as indicated in the RETURNED column.

Table 2 9103 Controller Communications Commands continued