8 Digital Communication Interface
Interface Commands
Command Description
Command
Format
Command
Example
Returned
Returned
Example
Acceptable
Values
Read R0 calibration parameter
r[0]
r
r0: 999.999
r0: 100.578
Set R0 calibration parameter to
n r[0]=n
r=100.324
90 to 110
Read ALPHA calibration
parameter
al[pha]
al
al: 9.9999999
al: 0.0038573
Set ALPHA calibration parameter
to
n
al[pha]=n
al=0.0038433
.002 to .005
Read DELTA calibration
parameter
de[lta]
de
de:9.99999
de: 1.507
Set DELTA calibration parameter de[lta]=n
de=1.3742
0–3.0
Read BETA calibration parameter be[ta]
be
be:9.999
be:0.342
Set BETA calibration parameter
be[ta]=n
be=0.342
–100.0 to 100.0
Functions not on menu
Read firmware version number
*ver[sion]
*ver
ver.9999,9.99
ver.9103,2.00
Read structure of all commands
h[elp]
h
list of commands
Read all operating parameters
all
all
list of parameters
Legend:
[] Optional Command data
{} Returns either information
n Numeric data supplied by user
9 Numeric data returned to user
x Character data returned to user
Note:
When DUPLEX is set to FULL and a command is sent to READ, the command is returned followed by a
carriage return and linefeed. Then the value is returned as indicated in the RETURNED column.
Table 2 9103 Controller Communications Commands continued