beautypg.com

Cautions about using the sensors – ThermoWorks TR-72-UI 2 CHANNEL LOGGERS WITH USB User Manual

Page 5

background image

vii

viii

Cautions about using the Sensors

Cautions about using the temperature sensor TR-0106

The possible temperature measurement range for
this sensor is
-40 to 110. Please use it within this
range.
It is possible to use just one extension cable per

temperature sensor.

Cautions about using

the temperature/humidity sensor TR-3100

The sensor can measure temperature within the
range of 0 to 50
and humidity within the range
of 10 to 95
RH. Only use the sensor within these
ranges.
If extremely severe temperature changes occur, the

humidity measurements may appear abnormal.

Once the sensor's temperature becomes stable, the

measurements will return to normal.
This sensor is not waterproof. Do not get wet.

The temperature/humidity sensor cable cannot be
extended.

When measuring with the temperature/humidity
sensor in an environment of 0 to 50
and less
than 30
RH, the measurements may sometimes
fluctuate. This is not abnormal.

Cautions about using the Sensors

[

Handling the temperature/humidity sensor]

-

The temperature/humidity sensor should probably be changed after a

period of about one year. After removing the sensor from its package,

please change the temperature/humidity sensor after one year’s usage.

During use the temperature/humidity sensor will accumulate impurities (dirt)

on the surface of the sensor causing the sensor's accuracy and sensitivity

to decrease. If the sensor is being used in a bad environment (smoky or

dusty places) it may be necessary to change the sensor sooner.

-

When the temperature/humidity sensor is not being used, please place it in

the attached vinyl bag with the drying agent included and store it in a cool

dark place with a temperature of between 5 to 25℃ and a humidity of below

30% RH.

-

Attached to the temperature/humidity sensor are two stickers: a wetness

detection sticker and a temperature detection sticker. If either of the stickers

shows abnormality, you should change the old sensor to a new one

immediately.

Wetness Detection

Sticker

Temperature Detection

Sticker

-

Wetness Detection Sticker

Informs you that the sensor has been wet.

Normal:

Under normal conditions, black

dots will appear on a white

background.

Abnormal:

Under abnormal conditions, it

will turn to red.

-

Temperature Detection Sticker

Informs you that the sensor measured a temperature measurement over 60℃ .

60

Normal:

Under normal conditions, the
number 「60」 will lightly appear on a
pinkish white background.

60

Abnormal:

Under abnormal conditions, the
number 「60」 will clearly appear on a
red background.

This manual is related to the following products: