background image

Measurement Probe selection LC-Display

Suitable for

(Prob)

these types

of measuring

instruments:

Temperature

Cu-CuNi Typ T

for all P700-

instruments

Humidity

%rF

P750/P755/P770

P755-LOG/P770-LOG

Flow

m/s

P750/P755/P770

P755-LOG/P770-LOG

Pressure

Pa

P750/P755/P770

P755-LOG/P770-LOG

Thermal flow

m/s

P750/P755/P770

probe

P755-LOG/P770-LOG

Note: Please check the probe selection to be sure
that the correct probe is entered. If a measurement
value is changed with the ENTER-key at the menu
Prob, the standard calibration will be automatically
used.

Indications for the combination probes
(temperature and humidity):
Please adjust relatively humidity as measurement
value for the channel on which the combination
probe is plugged. If you want that the measured
temperature of the plugged combination probe is
shown as well, you have to deactivate the channel
on which no probe is plugged on (see chapter
1.4.6./Chnl).

1.4.3 Difference temperature

(only 2-channel instruments) / [Lin2]

To display the difference temperature push
[ENTER/MENUE]. Use the up and down keys [

ᮡᮢ

]

to select Lin2. Push again [ENTER/MENUE]. On the
right corner of the display appears a T

1

-T

2

. Use the

up and down keys [

ᮡᮢ

] to adjust the requested sel-

ection. Push [ENTER/MENUE] to confirm. Push
[ESC] to be back in the measuring mode.

Note: Both channel have to be activated for showing
the difference temperature.

1.4.4 Calibration function / [CAL]

Despite high quality manufacturing techniques, each
probe is slightly different from specified standards.
To eliminate inaccuracies caused by exchanging or
ageing of probes, the instrument offer easy calibra-
tion functions which guarantee that the system
accuracy is always as good as if the instrument was
specifically calibrated to the individual probes in our
laboratory.

Important!

Important!

Important!

7

Unit

MENUE

Prob

Lin2

CAL

PrEn Chnl ArEA

Lo6

DE P700 Anleitung engl. 08/10 27.08.2010 14:02 Uhr Seite 7