beautypg.com

ThermoWorks P770 PRINTING DATALOGGER THERMOMETER User Manual

Page 18

background image

P750 (1-channel) / P755 (2-channel)

Inputs 1/2

Pt100,
thermocouple:
type K, J, L, N, R, S, T,
humidity, flow
resistance, voltage

Measuring range
Pt100

-200...+850 °C

Thermocouple

according DIN (-200...1760°C)

Resistance

0 ... 400

Humidity

0 %...100 %rH

Flow

0 ... 40 m/s

Accuracy
Pt100

±0.03°C from -50°C...+199.99°C
±0.05°C from -200°C...-50.01°C
otherwise 0.05%

Thermocouple
R, S

±1.0 °C +0.1%

K, J, L, N, T

±0.2 °C from 0 °C...+200 °C
±0.5 °C to 1000 °C
±1.0°C remaining range

Humidity

±1.5%rH

Flow

1% of end value

Resistance

0.5%

Resolution

0.0 1°C from -200 °C...+200 °C,
otherwise 0.1°C,
respectively 0.1%

Ex-mark

–––

Battery life

appr. 20 hours

Connectors

DIN 8-pole

Allowable
operating temp.

0 °C ... +40 °C

Display

2-line LCD

Housing

plastic

Dimensions

200 x 93 x 44 mm (LxWxH)

Weight

350 g

Power supply

9 V battery

18