USER MAINTENANCE

Warning

: Prior to any maintenance or service on this unit, disconnect the power feed from the power

supply.

Avertissement

: Avant d'effectuer toute maintenance ou entretien de cet appareil, débrancher le

cordon secteur de la source d'alimentation.

If a hazardous material/substance has spilled in the oven

, immediately initiate your site’s

Hazardous Material Spill Containment protocol. Contact your local Site Safety Officer and follow
instructions per the site policy and procedures.

C

LEANING

Note: The oven

’s chamber should be cleaned and disinfected prior to use.

Perform the steps below to clean the oven:

6. Remove the shelves and sliders.

7. Clean the oven with a mild soap and water solution, including all corners. Do not use an

abrasive cleaner that will damage stainless steel surfaces.

8. Rinse with distilled water and wipe dry with a soft cloth. Do not use deionized water.

9. Take care not to damage the temperature sensor probes when cleaning. These are located

on the right side of the of the oven chamber back wall.

Warning

: Never clean the oven with alcohol or flammable cleaners.

Avertissement:

Ne jamais nettoyer l'appareil à l'alcool ou avec des nettoyants inflammables.

Note: Do not use spray cleaners or disinfectants that might leak through openings and cracks and

get on electrical components, or that contain solvents that will harm coatings. Do not use
chlorine-based bleaches or abrasives; they will damage the

chamber’s interior surfaces.

D

ISINFECTING

Disinfect the oven on a regular basis if required by your production or laboratory protocol, or if mold
or fungus is an issue in your region. Perform the below steps to disinfect the oven:

1. Remove the shelves and sliders.

2. Disinfect the oven, including all corners and the access port, a using commercially available

disinfectant that is non-corrosive, non-abrasive, and suitable for use on stainless steel
surfaces. Contact your local Site Safety Officer for detailed information for the proper
disinfectants compatible with your application or process.

3. Take special care when cleaning around sensing heads to prevent damage and around the

door gasket so as not to impair the positive seal.