Operation (continued) – Shellab SMO5HP-2 User Manual
Page 21
![background image](/manuals/710840/21/background.png)
21 |
P a g e
OPERATION (CONTINUED)
Activating and Deactivating the Power Exhaust
To activate the power exhaust as part of a heating profile step, do the following:
1. While selecting the parameters
for the step, choose Event 1 On (“Ent1” “on”).
a. The Event 1 parameter is located after the Time parameter when scrolling
through a step menu using the green Advance Key.
b.
Use either of the Arrow Keys to toggle “oFF” to “on”.
2. Press the Advance key to scroll to the next event and save Event 1 as On.
3. Next time the profile runs, the 230 volt outlet on the back of the oven will power up
when the controller reaches this step.
4. The plugged in power exhaust will run until the next step in which Event 1 Off is selected
(“Ent1” “oFF”), or until the oven shuts down.
5. By default the Event 1 parameter is set to Off. It must be switched to On for each
successive step that you want the blower to run during the execution of the profile