Operation manual
SEA Series Steam Sterilizers
4
Prepare the machine before use
We have installed the
in the chamber(Fig3-3).
You should clean the
when use the autoclave
for a long time.The instruction shows you how to clean it.
Please refer to P17.
draining filter
draining filter
Before using the device, you should make electrical
connections by using the power cord that comes with
the unit. Insert the hollowed end of the power cord into
inlet of the machine and plug the other end on a power
outlet. Caution: The power supply can afford 1800VA load.
Ok, you can turn on the autoclave now.
Fig 3-3
Fig 3-4
The power supply must afford at least 1800VA load.
Tips: Before using the Autoclave, please check on the local atmospheric pressure. If the
pressure is lower than 0.095 Mpa or the Altitude is higher than 500 meters, you will
need to contact your distributor who should reconfigure the controller in the device.
Caution:
- AA13 (7 pages)
- GW1060 (44 pages)
- SS1 (4 pages)
- AA-8BS (30 pages)
- AA-22BS (32 pages)
- AA-8S (33 pages)
- AA-22S (26 pages)
- Shinva-45B (21 pages)
- AA9 (6 pages)
- USC-5L (5 pages)
- WS1.0 (3 pages)
- BEM4 (4 pages)
- LHF-40 (5 pages)
- NBCT2 (4 pages)
- NBCT7S (4 pages)
- ICB-12 (3 pages)
- LBC-4 (4 pages)
- SB-SDC-6 (5 pages)
- J-NBT (16 pages)
- BOB1 (5 pages)
- M-SOD-3636 (3 pages)
- J-SWB35 (21 pages)
- FXP16M (13 pages)
- M-SI-100 (3 pages)
- M-TMO1600 (3 pages)
- M-HOG-020 (13 pages)
- M-HOG010 (12 pages)
- J-USRC (16 pages)
- Z-L-450 (17 pages)
- Z-L-450A (6 pages)
- Z-L-530 (20 pages)
- M-SCEN-206 (13 pages)
- M-SCEN-207 (12 pages)
- Z-ZHWY-2102 (9 pages)
- GLF-90 (2 pages)
- PLCF300 (2 pages)
- H-DW-HW50 (12 pages)
- DW-FL90 (8 pages)
- B-IM0070FSC (5 pages)
- GLR-70 (3 pages)
- PLRGD-1400PT (3 pages)
- PVR-130 (3 pages)
- SPR-100 (3 pages)
- CEMLS (2 pages)