Т³гж 28, Form 3 error code chart, Operation manual – LABEC AA-22N User Manual
Page 27

Form 3 ERROR CODE CHART
Operation manual
SEA Series Steam Sterilizers
23
1
2
3
4
5
6
7
8
9
Serial number
E1
E2
E3
E4
E5
E6
E7
E8
E9
Code
Sensor error
Examine sensor
Examine inner temperature sensor
Super pressure(>2.6 Bar)
o
Chamber temperature(>145 C)
Examine outer temperature sensor
Door is opened when on working
Examine door switch
Pre-heating failed(>20 min)
Examine the electric connection of
temperature sensor, heat ring and heat stick
Error
Solution
Rerun parameter autoadjust program
Admission defeat(>30 min)
o
Steam generator ultra warm(>240 C)
Puts the steam defeat(20S<0.3Mba)
Inspect puts the steam valve, the vacuum valve
Inspect steam generator temperature sensor
Examine electric connection of heat stick, Admission pipeline
Constant temperature and
constant pressure failed
See also other documents in the category LABEC Equipment:
- AA13 (7 pages)
- GW1060 (44 pages)
- SS1 (4 pages)
- AA-8BS (30 pages)
- AA-22BS (32 pages)
- AA-8S (33 pages)
- AA-22S (26 pages)
- Shinva-45B (21 pages)
- AA9 (6 pages)
- USC-5L (5 pages)
- WS1.0 (3 pages)
- BEM4 (4 pages)
- LHF-40 (5 pages)
- NBCT2 (4 pages)
- NBCT7S (4 pages)
- ICB-12 (3 pages)
- LBC-4 (4 pages)
- SB-SDC-6 (5 pages)
- J-NBT (16 pages)
- BOB1 (5 pages)
- M-SOD-3636 (3 pages)
- J-SWB35 (21 pages)
- FXP16M (13 pages)
- M-SI-100 (3 pages)
- M-TMO1600 (3 pages)
- M-HOG-020 (13 pages)
- M-HOG010 (12 pages)
- J-USRC (16 pages)
- Z-L-450 (17 pages)
- Z-L-450A (6 pages)
- Z-L-530 (20 pages)
- M-SCEN-206 (13 pages)
- M-SCEN-207 (12 pages)
- Z-ZHWY-2102 (9 pages)
- GLF-90 (2 pages)
- PLCF300 (2 pages)
- H-DW-HW50 (12 pages)
- DW-FL90 (8 pages)
- B-IM0070FSC (5 pages)
- GLR-70 (3 pages)
- PLRGD-1400PT (3 pages)
- PVR-130 (3 pages)
- SPR-100 (3 pages)
- CEMLS (2 pages)