Т³гж 32, Troubleshooting – LABEC AA-22N User Manual
Page 31
![background image](/manuals/697844/31/background.png)
Operation manual
SEA Series Steam Sterilizers
27
Troubleshooting
1.During venting or drying, the thump sounds. It means that the condensed water collector is
full. You must drain the water from The collector.
2.Hissing from the chamber door. (Solving this problem according to item 8 Door Adjustment)
3.During a cycle, a banging sounds and large steam leaks from the chamber door and displays
E6
code. Please verify whether the door knob is properly turned and locked. If the steam
leaks at the beginning of a cycle, you need to change the seal according to instructions at page 18.
4.When a cycle completes, there are water drop on the instruments.
You may correct the problem by:
1). Please check the drain filter.
2). Reduce the load of the instrument in the chamber, especially the porous and dressing and
other water absorbent materials.
3). You may also wait for a while at the end of a cycle
Ed
for better drying result. The
autoclave will continue heating the chamber until you open the door.