8. Error messages
Error
message
Description Possible
causes
--ol-
Maximum load exceeded
Ö
Reduce load
Ö
Check whether the balance has
been damaged
Err 5
Keyboard error
Ö
Improper operation of the balance
Err 6
Value outside the A/D changer
range
Ö
Weighing plate not installed
Ö
Damaged weighing cell
Ö
Damaged electronics
Ba lo
Capacity of rechargeable
battery exhausted
Ö
Recharge battery
Should other error messages occur, switch balance off and then on again. If the error