KERN & SOHN HFM 10T1 User Manual
Page 22

Overview:
Function
Available
settings
Description
Low
High*
cap
d (low)
d (high)
1 t
2kg
1kg
500g 500g 200g 100g
3
t 10kg 5kg 2kg 2kg 1kg 500g
5
t 10kg 5kg 2kg 2kg 1kg 500g
10
t 20kg 10kg 5kg 5kg 2kg 1kg
F0 di
Modification of readability
AZn 0
0.5 d
AZn 1*
1 d
AZn 2
2 d
F1 az
Automatic zero point
correction
(zero tracking)
AZn 3
4 d
F2 bt
Not documented
F3 sp
Not documented
F4 ip
Internal A/D converter value
F5 ut
Not documented
Modifications may only be carried out by a
specialist with competent knowledge.
SLP 0
Standby mode switched off
SLP 1*
Standby mode after 5 minutes
SLP 2
Standby mode after 10 minutes
SLP 3
Standby mode after 20 minutes
F6 sl
Standby mode
see chap. 4.3.1
SLP 4
Standby mode after 30 minutes
F7 gv
Not documented
F8 ci
Password entry
In „P1- - - “ display increase the flashing number using
, select number using HOLD. Confirm entry by
ZERO button.
F9 CL
Adjustment, see chapter 7
* = default setting
HFM-BA-e-1011
22