KERN & SOHN KFS-TM User Manual
Page 54

Multi-interval balance
Confirm by
, after that the following menu items are
available.
Position decimal point available
selection 0, 0.0, 0.00, 0.000, 0.0000
inC
div 1
inC 1
Readability for
1. Weighing range
Selectable 1, 2, 5, 10, 20,
50
inC 2
�
inC 5
�
inC 10
�
inC 20
�
inC 50
�
div 2
inC 1
�
Readability for
2. Weighing range
Selectable 1, 2, 5, 10, 20,
50
inC 2
�
inC 5
�
inC 10
�
inC 20
�
inC 50
�
CAP
CAP 1
Balance capacity (Max) 1st weighing
range
CAP 2
Balance capacity (Max) 2nd weighing
range
Adjust weighing system after configuration.
CAL
nonLin
Adjustment, see chap. 6.5
LinEAr
For linearisation see chapter 6.6
F3 APP
Press adjustment switch
on
In verified weighing systems the access to the
configuration menu is locked.
off
Access to configuration menu enabled
(systems not appropriate for verification)
54
KFS-TM-BA_IA-e-1410