KERN & SOHN KFS-TM User Manual
Page 42
Overview:
Menu item
Available settings
F0 SEL
Enable tolerance check
1 SEL0
Tolerance control disabled
1 SEL1
Tolerance control for weighing
1 SEL2*
Tolerance control for counting
F1 Co
Display conditions of the tolerance
marker
11 Co0
Tolerance marker is always displayed, even if
standstill control is not yet displayed.
11 Co 1*
Tolerance marker is only displayed in connection
with standstill control.
F2 Li
Tolerance range
12 Li 0
Tolerance marker is only displayed above zero
range.
12 Li 1*
Tolerance marker is displayed for the whole
range.
F3 Pn
Number of limiting points
13 Pn 0
1- Limiting point (OK/ -)
13 Pn 1*
2- Limiting points (+/OK/-)
F4 bU
Audio signal
14 bu0*
Audio sound during tolerance control disabled
14 bu1
Audio sound when load is within tolerance limits
14 bu2
Audio sound when load is beyond tolerance limits
F5 Ao
Automatic zero point correction
(zero tracking)
2 Ao0
Automatic zero tracking off
2 Ao1
Automatic limiting point correction on, 0.5 d
2 Ao2*
Automatic limiting point correction on, 1 d
2 Ao3
Automatic limiting point correction on, 2 d
2 Ao4
Automatic limiting point correction on, 4 d
F6 At
Automatic cutout
on
Automatic shut-off ON
off
Automatic shut-off OFF
F7 AP
Automatic shutdown for battery
operation
3 Ap0*
AUTO OFF function disabled
3 Ap1
Instrument will be switched off after 3 minutes of
inactivity of display unit or weighing bridge.
F8 UA
RS-232 mode
4 UA0
Output via RS232C interface disabled
4 UA1*
Continuous data output
4 UA2
Continuous data output of stable weighing values
4 UA3
One output for stable weighing value. No output
for stable weighing values. Renewed output after
stabilization.
4 UA4
For remote commands,
see chap. 9.2.
Issue after pressing the PRINT ´button
4 UA5
Standard printer setting, output after pressing the
PRINT button
id on/off
Printout memory on/off
dt on/off
Printout date on/off
G on/off
Printout gross weight on/off
n on/off
Printout net weight on/off
C on/off
Printout total on/off
PCS on(off)
Printout parts counting on/off
4 UA6
Not documented
42