9 rs 232c interface, 1 technical data – KERN & SOHN KFS-TM User Manual
Page 44
9 RS 232C interface
You can print weighing data automatically via the RS 232C interface or manually by
pressing
via the interface according to the setting in the menu.
This data exchange is asynchronous using ASCII - Code.
The following conditions must be met to provide successful communication between
the weighing system and the printer.
• Use a suitable cable to connect the display unit to the interface of the printer.
Faultless operation requires an adequate KERN interface cable.
• Communication parameters (baud rate, bits and parity) of display unit and
printer must match.
9.1 Technical data
Pin No.
Signal
Input/output
Functions
1
EXT.TARE
Input
External tare
※2
➁
2
RXD
Input
Receiving data
3
TXD
Output
Transmission data
4
DTR
Output
HIGH level with power “on”
5
GND
―
Signal ground
6
―
―
―
7
―
―
―
8
―
―
―
9
GND
―
Signal ground
1 2 3 4 5
6 7 8 9
9 pin d-subminiature bushing
44
KFS-TM-BA_IA-e-1410