7.5 Weighing with tare
Deposit weighing vessel. After successful standstill control press the
button.
Zero display and the indicator [
] next to
T AR E
appear. The weight of the
container is now internally saved.
Weigh the material, the net weight will be indicated.
After removing the weighing container, the weight of the weighing container
appears as negative display.
The tare procedure can be repeated as many times as necessary, for example
with initial weighing of several components for a mix (add-on weighing). The limit
is reached when the total weighing range capacity is full.
To delete the tare value, remove load from weighing plate and press
.
7.6 Counting
During piece counting parts can either be counted into a container or out of a
container. To count a greater number of parts the average weight per part has to be
determined with a small quantity (reference quantity). The larger the reference
quantity, the higher the counting exactness.
High reference must be selected for small parts or parts with considerably different
sizes.
The average piece weight can only be determined by stable weighing
values.
If weighing values are under zero, the piece counter display shows a
negative number of items.
The message
LIGHT
appearing on the display indicates that load falls
below minimum weight value.
Delete incorrect entries by pressing
.
The accuracy of an average item weight can be improved at any time
during additional counting processes. For this purpose add additional
items and press
. After the reference optimization sounds a signal
tone. As the additional pieces increase the base for the calculation, the
reference also becomes more exact.
22
KFS-TM-BA_IA-e-1410