Configuration menu overview:
Menu block
Main menu
Menu item
sub menu
Available settings / explanation
F0 iSn
-
Display internal resolution
F 1 Grv
-
Not documented
F2 dm
Single-range balance
Confirm by pressing
, then the following menu items can
be selected by
.
dESC
Position decimal point available selection
0, 0.0, 0.00, 0.000, 0.0000, 0.00000
inC
inC 1
Readability
selectable 1, 2, 5, 10, 20, 50
inC 2
inC 5
inC 10
inC 20
inC 50
CAP
Balance capacity (max)
Adjust weighing system after configuration.
CAL nonLin
Adjustment, see chap. 6.5
LinEAr
For linearisation see chapter 6.6