2 remote control instructions – KERN & SOHN AEJ 200-4CM User Manual
Page 94
![background image](/manuals/697267/94/background.png)
16.4.2 Remote control instructions
Feedback by weighing scale after sending remote control command:
XX_A
CR LF
Instruction accepted; will be executed
XX_I
CR LF
Instruction received; impossible to carry out
XX_^
CR LF
Command accepted but max range exceeded
XX_v
CR LF
Command accepted but min range exceeded
XX_E
CR LF
Error during execution, timeout for stable weighing value
exceeded
XX
Name of command
1 – 3
4
5
6
7
8 – 16
17
18 - 20
21
22
C
om
m
a
nd
B
lank
S
ta
b
ili
ty
s
y
m
bol
B
lank
C
har
ac
ter
W
ei
gh
t
B
lank
Un
it
CR
LF
Instruction:
1-3
Character
Stabilisation
symbol
[blank space]
If stable
[?]
If stable
[^
]
If error (Command + exceeded)
[v]
If error (Command – exceeded)
Character
[blank space]
For positive and negative values
Weight
9 digits
With right justifications
Unit
3 digits
With left justification
94
AEJ-C/AES-C/PLJ-C-BA-e-1412