2 menu settings – KERN & SOHN AEJ 200-4CM User Manual
Page 27
7.2 Menu settings
Menu item „P1 ADJUSTMENT“
[Adjustment]
AEJ-C / PLJ-C
AES-C
P1.1 INT.JUST.
-
Internal adjustment, ILJ-C only
-
P1.2
EXT.JUST.
External adjustment *
-
P1.3
BE.JUST.
User adjustment
P1.4 TEST JUST.
-
Adjustment test
* = Locked in weighing scales with type approval certificate [Er 9 lock].
7.2.1 P1.1 INT.J US T. - internal adjus tment (only models AE J -C and PLJ -C)
• For automatic internal adjustment see chap. 7.1
• Starting internal adjustment manually
Press the CAL-key or invoke menu “P1.1 INT.JUST.“
Press
or
Internal adjustment will be taking place.
• Ensure that there are no objects on the weighing pan.
• You can cancel the adjustment by pressing the UNIT-key.
• After successful adjustment the weighing scale will return automatically to
the menu item. To return to weighing mode, press UNIT.
AEJ-C/AES-C/PLJ-C-BA-e-1412
27