13.4 Percent determination

Percent determination allows weight display in percent, in relation to a reference
weight. The reference weight may be determined by weighing or by entering it
numerically.
Determining reference by weighing

The function DETERMINE REFERENCE WEIGHT is allocated to the F1-key.

Go to menu and select P2.6 DEVIATION

Use the arrow keys to select P2.6.5 SHORTCUTS

Press F2, P2.6.5.1 will be displayed along with the previous function for the
F1-key.

Again press F2 and the information line will start flashing.

Use the arrow keys to select DETERMINE REFERENCE WEIGHT and

Confirm by PRINT; this function has now been allocated to F1.


In weighing mode:

This completes setting the operating
mode weighing by percentage.

The percent symbol will be
displayed at the upper margin of the
display.
The setting SAMPLE 0.0 G will be
displayed at the lower margin.

Put the reference weight equalling
100% on the weighing plate.

(example)

Example: 1 kg is supposed to be
equivalent to 100%.

The reference weight of 1 kg has now been saved
as 100% value.

Remove reference weight and place load. The weight will be displayed in
percent according to reference weight.