Appendix 5: parameter’s default table, Ppendix, Arameter – Flintec FT-16D V.1.21 User Manual
Page 31: Efault, Able
A
PPENDIX
5: P
ARAMETER
’
S
D
EFAULT
T
ABLE
FT-
Parameter Factory
defaults
16 16D
0-- Interface
x
x
01- Serial Interface 2
x x
10 Data format
0 = Disabled
x
x
11 Baud rate
3 = 9600
x
x
12 Handshake
0 = No
x
x
13 Address
0
x
x
14 Data length and parity
0 = 8 bit, no parity
x
x
02- Serial Interface 3
x x
20 Data format
0 = Disabled
x
x
21 Baud rate
3 = 9600
x
x
22 Handshake
0 = Disabled
x
x
23 Address
0
x
x
24 Mode
0 = RS232C
x
x
25 Data length and parity
0 = 8 bit, no parity
x
x
04- Printer
x x
45 Minimum print
20 kg
x
x
46 Print control
0 = With Print Key
x
x
1-- Configuration
x
x
11- Start
Up
x x
114 Buzzer active?
0 = Disabled
x
x
12- Filter
x x
120 Filter
FT-16 : 7 / FT-16D : 3 x
x
2-- Scale Parameters
x
x
20- Set
up
x x
200 Approved
0 = No
x
x
201 Increased indication
0 = With x10 key
x
x
202 Power on zero
2 = ± %10
x
x
203 Zero with key
1 = ± %2
x
x
204 Auto zero tracking
0 = Disabled
x
x
205 Tare
1 = With Tare Key
x
x
206 Motion detector
1 = ± 0,5e
x
x
21- Scale
Build
x x
210 Scale Type
0 = Single Range
x
x
211 Capacity 1 / d1
x
x
212 Capacity (2) / d(2)
60 t / 20 kg
x
x
22- DLC Configuration Block
x
220 DLC Type
0 = Flintec RC3D
x
221 Quantity of DLC
1
x
222 DLC Address
x
3-- Calibration / Adjustment
x
x
30- Calibration
x x
300 Gravity
x
301 Calibration
x
x
31- Adjustment
x x
310 Zero adjustment
x
x
311 Span adjustment
x
x
312 Span adjustment under load
x
32- Eccentricity
Adjustment
x
320 Adjustment Method
0 = independend DLC
x
321 Auto Adjustment
x
322 Manual Adjustment
x
323 Temporarily set constants to “1” 0 = normal operation
x
8-- Metrological Data
x
x
80- Legal
Metrology
x x
800 Calibration counter
x
x
801 Alibi memory
x
x
802 Calling a record
x
x
803 Print Alibi memory
x
x
804 Alibi memory information
x
x
FT-
Parameter
16 16D
9-- Diagnostics
x
x
90- Tests
x x
900Key Pad test
x
x
904mV indication
x
905DLC internal count
x
91- Serial
numbers
x x
910FT-1x serial no.
x
x
911Option 1 serial no. (if installed)
x
x
912Option 2 serial no. (if installed)
x
x
913Option 3 serial no. (if installed)
x
x
97- Calibration
coefficients
x
97X Calibration coefficients
x
99- Print
Parameters
x x
990All parameters
x
x
991Load default parameters
x
x
FT-16(D) Technical Manual, Rev. 1.21 December 2008