Appendix 4: error table, Ppendix, Rror – Flintec FT-16D V.1.21 User Manual
Page 30: Able
A
PPENDIX
4: E
RROR
T
ABLE
The FT-16(D) weight indicator has been designed as very reliable and virtually error free instrument. However if
an error occurs do not attempt to repair the equipment before you understand what caused the error. Note the
problems you have with your instrument and the error messages shown on the display. Then try to solve the
problem according to the error table given below.
Error
Code
Description
Actions to take / Possible cause
FT-16
FT-16
Under
Weight is too low
x
x
Over
Over Load
Check the load; Load cell or instrument could
be defective
x x
Adc Out
Input signal outside the input range
Check the load; Check the calibration; Load
cell or instrument could be defective
x
E E E
Load not within zero range
See parameter [202] in chapter 6.4
x
x
Err 1
x x
Err 2
ADC error respective
Load cell communication error (DLC)
Re-energize indicator; contact FLINTEC
x x
Err 3
Indicator cannot get calibrated
Check load cell and load; repeat the
calibration
x x
Err 10
EEPROM error
Configure the instrument; EEPROM defective
x
x
Err 20
Calibration error
Calibrate the indicator
x
x
Err 21
Configuration error
Configure the indicator
x
x
Err 22
Tare, CN, Total weight and the SP,
PT and ID in use error
Check SP, PT and ID entries; Check Tare,
CN and Total weight
x x
Err 27
Indicator is not calibrated
Calibrate the indicator
x
x
Err 30
Processor Error
Contact FLINTEC
x
x
Err 34
Indicator cannot get calibrated
Load cell signal is negative , very low or too
high
x x
Err 35
Calibration Error
Calibration test weight is too small; Check
load cell connections
x x
Err 37
Scale unstable
Wait until scale becomes stable; Check
ground connection
x x
Err 40
Alibi memory error
Correct parameter [801]; Alibi memory card
broken
x x
Err 41
New alibi card installed
x
x
Err 42
Alibi card serial no. is not identical
Appears when new alibi memory card is
installed (*)
x x
Err 43
Alibi card serial no. is not identical, no
print operation will be performed
Identify the alibi memory card to the indicator
(*)
x x
Err 44
Alibi card info error
x
x
Err 46
Alibi recording capacity error
x
x
Err 47
Info error from weighing board
Contact FLINTEC
x x
Err 61
U10 (eeprom) is not installed
Contact FLINTEC
x
x
E XXXX
Hardware error
Contact FLINTEC
x
x
Ed1 yy ** Communication time out
x
Ed2 yy ** Checksum error
Check DLC cables; Check DLCs and serial
number; contact FLINTEC
x
Ed3 yy ** DLC error
Check ambient temperature; replace DLC
x
Ed4 yy ** DLC is under
x
Ed5 yy ** DLC is Over
Check mechanical installation and DLC
x
Ed6 yy ** Interface board error
Replace main board or DLC board
x
Ed7 yy ** DLC addressing error
Check serial number and re-define DLC
address OR replace DLC
x
Ed0 yy ** Other DLC error
Replace DLC
x
** yy = number of DLC
FT-16(D) Technical Manual, Rev. 1.21 December 2008
Page 30 of 32