9 diagnostics, Diagnostics, ] diagnostics – Flintec FT-16D V.1.21 User Manual
Page 23: ] tests, ] serial numbers, ] calibration coefficients (for ft-16 only), ] printing parameter values
9 D
IAGNOSTICS
The parameter block [9--] contains certain functions for checking and testing.
[9--] Diagnostics
The operations about checking and testing the instrument can be made here.
[90-] Tests
[900 ] Key Pad test
In this step each key’s ASCII code will be shown on the display as you press the related key. By this way
you can test if all the keys are functioning or not. Pressing “
↑” key will take you to the next parameter.
[904 ] mV Indication (for FT-16 only)
When pressing the
uncalibrated value is only for test- and service purposes.
[905 ] DLC Internal Count (for FT-16D only)
When pressing the
selected by up and down keys and internal counts can be seen by pressing the
Press the
[91-] Serial Numbers
[910]
Instrument Serial No.
[911]
Option 1 Serial No.
[912]
Option 2 Serial No.
[913]
Option 3 Serial No
When pressing the
[97-] Calibration Coefficients (for FT-16 only)
The parameters located in this section are the calibration parameters of the instrument. If these values have
been noted this can help when a calibration problem occurs. You can refresh the calibration by entering the
correct values.
[99-] Printing Parameter Values
[990] Print All Parameters
By pressing the
[991] Loading Default Parameters
When pressing the
loading the default values or press the