beautypg.com

Configuration parameters, Scale parameters, 3 configuration parameters – Flintec FT-13 User Manual

Page 15: 4 scale parameters, ] configuration block [12-] filter, ] entries, ] scale block, ] basic set up, ] scale build

background image

FT-13 Technical Manual, Rev. 1.40 March 2013

Page 15 of 56

6.3 Configuration Parameters

[1--] Configuration Block

[12-] Filter

In this block the proper filter values according to the operating conditions can be entered. One of the most
important features of the indicator is viewing the filter characteristics on the display and with the help of this
option, you can select the most suitable filter without leaving the setup mode.

[120 X ] Filter

The filter value can be set from 0 to 9 (The minimum value of the filter is at least 6 in normal weighing applications).
When you enter this parameter and press the key while [120 X] is shown on the display, the weight
variation can be seen on the display. The value of the filter can be changed by using the < > and < > keys and the
weight variation for each value is shown on the display. After finalising the filter selection you can go to the next step
by pressing the key.

[14-] Entries

[140 ] Date

[DD.MM.YY]

First press a key for the date entry, then enter the new date via numerical keys and save by
pressing .

[141 ] Time

[ HH.MM]

First press a key for the time entry, then enter the new date via numerical keys and save by
pressing .

6.4 Scale Parameters

[2--] Scale Block

The parameters related to the weighing operation are located here.

[20-] Basic Set Up

[200 X] Approved

0 = No

1 = Yes

[201 X] High Resolution Indication

0 = by pressing key

1 = Always increased indication

[203 X] Zeroing Range

0 = disable

1 = ± 2%

2 = ± 20%

[204 X] Auto Zero Tracking

For compensating selected small deviations around the centre of zero this function automatically re-
adjusts the scale to zero.

0 = disable

1 = ± 0.1e

2 = ± 0.2e

3 = ± 0.3e

4 = ± 0.5e

5 = ± 1e

[205 X] Tare

0 = Disable Tare pushbutton

1 = Enable Tare pushbutton (in weighing mode)

[206 X] Motion Detector

This parameter defines the sensitivity level which will determine what is considered as stable.

0 = ± 0.3e

1= ± 0.5e

2= ± 1e

3= ± 2e

4 = No motion detector

[21-] Scale Build

The capacity and the resolution of the scale will be defined here.

[212 ] Capacity

Press the key to reach this parameter.

[CAP ]

The capacity of the scale will be entered here by the key and the key.
The value will be confirmed by pressing the key.

[ XXXXXX ]
[d ]

Display resolution of the scale will be selected by the key.
The selection will be confirmed by pressing the key.

[ XXXXXX ]