Culligan Drinking Water System User Manual
Page 10
If you wish to service your system, the filters are available from your Culligan dealer. To locate your
nearest Culligan dealer, call (800) 285-5442, or visit our website at www.culligan.com. See the Parts
List at the end of this section for replacement numbers.
Cartridge Conditioning
The activated carbon, reverse osmosis, and polishing filter cartridges must be conditioned as follows
prior to installation into the Culligan Water Tower™ drinking water system. Your Culligan Man™ can
do these important procedures at the time of purchase.
Activated Pre-Carbon Filter Cartridge - 2 gallon flush to remove carbon dust
RO Membrane Cartridge - 24 hour flush to remove preservative solution
Activated Post-Carbon (Polishing) Filter Cartridge - 2 gallon flush to remove carbon dust
Tubing Connectors
The Culligan Water Tower drinking water system features easy-to-use push-in tubing connectors
(Figure 3). Fittings consist of two parts: a body and a colored collet. Collet color corresponds to tub-
ing color to be used at that connection (Figure 3A).
1. To install a tube, lubricate the tubing end with water (only) and push it through collet until it
seats firmly at bottom of fitting (Figure 3B). You should feel the tubing pass the O-rings. Cut
ends must be square. Be sure the tubing is smooth and free of abrasions, otherwise a leak may
result. Cut tubing only with a new razor blade.
2. To remove a tube, push and hold the collet against the fitting body and pull the tube from the
fitting. (Figure 3C).
Figure 3A
Figure 3A
Figure 3A
Cartridge Replacement, Cleaning and Sanitizing Procedure
1. Mix a mild cleaning solution of dish soap and clean potable water in plastic bowl.
2. Empty storage tank and relieve system pressure. Verify tank valve is open. Close feed water sup-
ply valve and open product water faucet.
Note: Additional point-of-use devices (i.e. icemakers) may use filters along their supply line.
Remove any filter or treatment device installed between module and delivery device before pre -
ceding.
Icemaker: Transfer ice cubes from bin/tray to clean freezer container for storage until procedure
is done.
3. Check product water storage tank air pre-charge using low-pressure gauge (PN 34002024). Air
valve is located on tank base. Pre-charge should be 55 kPa (8 psig) with tank empty and tank
valve open.
10