beautypg.com

Culligan IC-100 User Manual

Disposable icemaker/water dispenser filter, Filter installation instructions, Use and care manual model ic-100

background image

Precautions and Usage Guidelines

WARNING:

Do not use with water that is microbiologically unsafe or of

unknown quality without adequate disinfection before or after the system.

CAUTION:

The filter must be protected against freezing.

Failure to do so may result in cracking of the filter and water leakage.
NOTE:

Make certain that installation of this system complies with all state and
local laws and regulations.

FOR COLD WATER USE ONLY.
MAINTENANCE:

Contact your area retailer for replacement cartridges and

pricing or call Culligan Customer Service at 800-634-1455 for the nearest
dealer. Change cartridge every 2500 gallons or approximately every 12 months.

Filter Installation Instructions

(For copper or plastic tubing)

NOTE:

Shut off water supply when replacing filter.

1. Shut off water supply. Select a location for installation on 1/4" OD (outside

diameter) plastic or copper tubing. Filter requires at least three

inches behind appliance. See Fig 1.

NOTE:

Leave at least a 5-foot coil of tubing between appliance side and

compression fittings. This will allow room to pull out the appliance for
changing the filter.

2.

Wrap 6" of Teflon tape around the threads of both ends of fittings.
Screw fittings into the body of the filter until hand tight. Use a wrench to
tighten two additional turns. DO NOT OVER-TIGHTEN. About one full thread
should remain visible.

3.

Two types of compression nuts are included: One for copper tubing and one
for plastic tubing. Copper tubing uses nuts with metal grippers, while the
plastic tubing uses the nuts with plastic grippers. Refer to Fig. 2.
DO NOT DISASSEMBLE NUTS. If nuts become disassembled, refer back to
Fig 2 for reassembly. Attach correct compression nuts to fittings.

4.

Cut 1/4" OD tubing where filter will be installed.

5.

On the inlet side of filter, insert tubing as far as possible into fitting. WATER
FLOWS IN DIRECTION INDICATED BY ARROW ON THE FILTER. See Fig 3.
If tubing does not enter the fitting, loosen the nut one turn. After inserting
tubing, hand tighten nut first, then tighten 1-1/2 to 2 turns with a wrench.

6.

With only the inlet side connected, flush water thru filter into a container
for 5 minutes. Wait an hour, and then flush another 5 minutes, or until water
runs clear.*

7.

Shut off water and then connect the outlet side of the filter to the appliance
using the procedure in step #5. Turn water on and check for leaks.

*NOTE:

Though the filter has been rinsed to remove fine particles of carbon,

resettling of the carbon within the filter can occur. If the first trays of ice
become discolored, discard the ice until clear.

Disposable Icemaker/Water Dispenser Filter

Use and Care Manual Model IC-100

QUESTIONS? Call Culligan Customer Service: 1-800-634-1455, Mon.-Fri., 7:30 AM-7 PM CST

145233 Rev D 03/04

Culligan

Customer Service M-F 7:30 a.m – 5:00 p.m. CST

Phone: 1-800-634-1455 Fax: 1-888-749-8826

International: Phone (920) 457-9435 • Fax (920) 457-2417

e-mail: [email protected]

Sales & Marketing Correspondence

P.O. Box 1086 • Sheboygan, WI 53082-1086

Fax (847) 291-7632

e-mail: [email protected]

©2004 Culligan International Company, Printed in U.S.A.

Culligan International Company

One Culligan Parkway

Northbrook, Illinois 60062

www.culligan.com

Figure 2

Figure 1

Figure 3

To Refrigerator

Supply Side

Plastic Gripper

Metal Gripper

Outlet

Inlet

Performance Data

Important Notice: Read this performance data and compare the capabilities of this system
with your actual water treatment needs. It is recommended that, before installing a water
treatment system, you have your water supply tested to determine your actual water treat-
ment needs.
This system has been tested according to NSF/ANSI Standard 42 for reduction of the
substances listed below. The concentration of the indicated substances in water entering the
system was reduced to a concentration of less than or equal to the permissible limit for
water leaving the system, as specified in NSF/ANSI standard 42.

Test Conditions:

Flow Rate

= 0.75 gpm (2.8 L/min)

Inlet Pressure = 60 psi (4.14 bar)
pH =

7.5

± 1

Temperature = 68º F ± 5º F

(20º C ± 2.5º C)

Operating Requirements:

Pressure =

30-125

psi

(2.07-8.62 bar)

Temperature =

40º-100º F
(4.4º-37.7º C)

Turbidity

= 5 NTU Max

Substance

Influent Challenge

Max. Permissible Product

Reduction

Minimum

Average

Concentration

Water Concentration

Requirements

Reduction Reduction

Standard 42
Chlorine

2.0 mg/L±10%

≥ 50%

>93%

Flow Rate=0.75 gpm (2.8 L/min) Capacity=2500 gallons (9463 L) or 12 months
Testing was performed under standard laboratory conditions, actual performance may vary.

The IC-100 is Tested and Certified by NSF International to NSF/ANSI Standard 42 for
the reduction of Chlorine, Taste and Odor.

IOWA RESIDENTS ONLY:

Store or seller’s name

Address

City

State

Zip

Telephone

Seller’s signature

Customer’s signature

Date