Step 2.
Use a megohmmeter and measure the
resistance between all five wires tied
together and the load cell metal body.
• The measured value should be 5,000
megohms or more.
WARNING
WHEN USING A MEGGER DO NOT EXCEED
50 VOLT RANGE.
• If the sensor fails this test remove the
ground wire and test with only the four
live leads.
• If the sensor passes the test an insulation
problem in the cable is most likely.
Step 3.
Replace the load cell if the cell fails both
tests.
Electrical Termination
Cable Color Codes
The cable is 6 conductor, shielded (floating) and 20
feet in length.
EXC+
Red
EXC -
Black
SHIELD
Yellow
C2+
Gray
C2-
Violet
SIG +
Green
SIG -
White