Hardy HI LPRE05 Shear Beams User Manual
Page 14
HI HLPS SERIES LOAD POINT ASSEMBLY
Page 8
Round Vessel with 3
Load Point
Assemblies - Angle
Config. #2
FIG. 8: ANGLE FOR ALL LOAD CELLS IN ROUND
CONFIGURATION
NOTE:
Load Cells can be oriented with cables pointed either
in or out. For three load point systems, the mounting
locations should be spaced120 degrees apart. For
four load point systems the mounting locations should
be spaced 90 degrees apart.
Square Hopper with
3 Load Point
Assemblies - Even
Load Distribution
FIG. 9: ANGLE FOR FIXED AND BUMPER LOAD
CELLS IN SQUARE CONFIGURATION