Chapter 5 - Block Transfers
5-19
Block Write
Command Number
54: Scale
Calibration Action
TABLE 5-21: BLOCK WRITE COMMAND NUMBER 54: SCALE CALIBRATION ACTION
BLOCK WRITE COMMAND NUMBER 54: Scale Calibration Action
START
WORD DEFINITIONS:
#WORDS
WORD
Command number: A value of 54 (decimal)
bit 0
0
bit 1
1
bit 2
1
bit 3
0
bit 4
1
bit 5
1
bit 6
0
bit 7
0
Remote Functions Byte
bit 8
Setting then clearing this bit tells the instrument that current
weight is an empty scale.
bit 9
Setting then clearing this bit tells the instrument that current
weight is span weight.
bit 10
Setting then clearing this bit stores critical data in the Secure
Memory Module.
bit 11
Setting then clearing this bit restores critical data from the
Secure Memory Module.
bit 12
Setting then clearing this bit tells the instrument that current
weight is Midpoint Linearity value.
bit 13
Reserved for future use
bit 14
Setting then clearing this bit tells the instrument that current
weight is the C2™ reference point.
bit 15
Reserved for future use
1
0
TOTAL NUMBER OF WORDS
1