rECommEndations and dECLaration By tHE manufaCturEr -
- ELECtromaGnEtiC immunity
The Compex Runner is designed for use in the electromagnetic environment
stipulated below. The buyer or user of the Compex Runner must ensure it is
used in this recommended environment.
The Compex Runner is designed for use in the electromagnetic environment
stipulated below. The buyer or user of the Compex Runner must ensure it is
used in this recommended environment.
Immunity test
Test level IEC 60601
Observance level
Electromagnetic
environment - recommendations
Electrostatic discharge (DES)
IEC 61000-4-2
± 6 kV at the contact
± 8 kV in air
± 6 kV at the contact
± 8 kV in air
Floors must be wood, concrete or ceramic tiles. If floors are covered with synthetic
material, the relative humidity must be maintained at a minimum of 30%.
Rapid electrical transients in
bursts of noise
IEC 61000-4-4
± 2 kV for power supply lines
± 1 kV for input/output lines
Not applicable
System battery-powered
The quality of the power supply should be that of a typical commercial or
hospital environment.
Surge currents
IEC 61000-4-5
± 1 kV differential mode
± 2 kV common mode
Not applicable
System battery-powered
The quality of the power supply should be that of a typical commercial or
hospital environment.
Voltage trough, short power
cuts and voltage variations
on power supply input lines
IEC 61000-4-11
< 5% VT (trough > 95% of
VT) for 0.5 cycle
< 40% VT (trough > 60% of
VT) for 5 cycles
< 70% VT (trough > 30% of
VT) for 25 cycles
< 5% VT (trough > 95% of
VT) for 5 seconds
Not applicable
System battery-powered
The quality of the power supply should be that of a typical commercial or hospital
environment. If the Compex Runner user requires continuous operation during mains
power cuts, we recommend that the Compex Runner is powered by a UPS or a battery.
Magnetic field at mains
frequency (50/60 Hz)
IEC 61000-4-8
3 A/m
Magnetic fields at the mains frequency should be at the level of a representative site
located in a typical commercial or hospital environment.
notE: Vt is the aC supply voltage before application of the test level.
notE: Vt is the aC supply voltage before application of the test level.