Page 13

Warnings and precautions
991-3001 Rev. 1.3
June-2013
xiii
Warnings and precautions
Whenever the
symbol appears on the NucleoCounter
®
NC-
3000™ instrument, it
indicates that the manual must be consulted for precautions and warnings.
Power and cables
Use the shielded USB cable supplied with the NucleoCounter
®
NC-
3000™ to ensure
that appropriate EMI classification is maintained for the intended environment.
The USB interface connector of the NucleoCounter
®
NC-
3000™ must only be
connected to SELV circuits. External computing devices connected to the USB
interface connector of the NucleoCounter
®
NC-
3000™ has to comply with the
standards, UL 1950 and IEC/EN 60950.
Table 1. Description of the NucleoCounter
®
USB interface connector
NucleoCounter
®
NC-
3000™ is powered by an external 24VDC power supply. For
safe use, please follow the instructions for connecting the power supply.
The NucleoCounter
®
NC-
3000™ shall only be used with an external power
supply purchased at ChemoMetec A/S.
The detachable DC power supply cord set and appliance inlet of the external
power supply are considered as the disconnecting device.
The mains supply cord and plug of the external power supply shall comply with
any national regulations.
1
In normal operation mode (refers to Pin no 4)
Pin no.
Name
Maximum
Voltage level
1
1
+5V
+5 VDC
2
D-
+3.5 VDC
3
D+
+3.5 VDC
4
DGND
0 VDC
Metal
enclosure
Shield
(connected
to DGND)
0 VDC