3.0

CALIBRATION CURVES


Type of wood

Filling quantity

Measuring range

Beech

50 g

15 % to 45%

Beech

100 g

40 % to 60 %

Birch

50 g

15 % to 45%

Birch

100 g

35 % to 55%

Oak

50 g

15% to 45%

Oak

100 g

40% to 55%

Spruce

50 g

15% to 60%

Spruce

100 g

50% to 65%

Poplar

50 g

15% to 60%

Poplar

100 g

50% to 60%

Reference

------

------


To verify the calibration curve only the standardized drying method is permissible
because many other methods used are doubtful and can lead to incorrect
comparison values. If no existing curve fit, then a calibration by the company
Schaller is required.




4.0

DETERMINATION OF THE MATERIAL REFERENCE MOISTURE



The principle is a comparison measurement with the dehydration method according
to CEN/TS 14961:2005. Take the measured sample and weigh it. Dry it out in an
oven and weigh it again.







100

%

×

=

Mn

Mt

Mn

F

M

n

: Mass with average moisture content

M

t

: Mass of the dried sample

%F: Calculated moisture content

Checkline-Europe

-11-

10.0

DEVICE MAINTENANCE INSTRUCTIONS


To provide a long life of your device please does not expose it to strong mechanical
loads or heat e.g. dropping it or direct sunlight exposure. Clean your device using a
dry cloth. Any kind of wet cleaning damages the device.

It is not allowed to load the measuring head incorrect (stress, bending), otherwise it
can be broken. Plug and remove the insertion probe of your BLH straight into the
wood chips.

The instrument is not rainproof. Keep it in dry areas. When the device isn’t used for
a longer period (2 months) or when the batteries are empty, they should be removed
to prevent a leakage of the battery acid.


11.0 EXEMPTION FROM LIABILITY

For miss readings and wrong measurements and of this resulting damage we refuse
any liability. This is a device for quick determination of moisture. The moisture
depends on multiple conditions and multiple materials. Therefore we recommend a
plausibility check of the measuring results. Each device includes a serial number
and the guarantee stamp. If those are broken, no claims for guarantee can be
made. In case of a faulty device, please contact Checkline Europe (www.checkline-
europe.eu) or our dealer.