12.0 TECHNICAL DATA
Resolution of the display
0.5% water content
0.5°C temperature
Measuring range
15 up to 65% depending on the material
Operation temperature
0°C up to +40°C
Storage temperature
-20°C to +60°C
Temperature compensation
Automatically
Power supply
4 pcs. 1.5 Volt AA Alkaline batteries (900
measurements)
Auto Switch OFF
After app. 6 minutes (adjustable)
Current consumption
60mA (with light)
Display
128 x 64 matrix display, lighted
Dimensions
275 x 180 x 290 mm
Weight
ca. 2,9 kg (including batteries)
Degree of protection
IP 40
Scope of supply
Device including measuring cup
4x1.5 Volt AA Alkaline batteries
Software LogMemorizer
USB interface for PC transfer
USB cable
Checkline-Europe
-5-
2.1 Information of the measurement of sawdust
Enclosed you will find instructions for measuring sawdust. If you are not sure what
type of wood you use, we recommend a one-time measurement compared with the
drying method according CEN / TS 14774.
The calibration curves using 100g filling quantity only be used in very wet material.
In general, the sample size of 50gr has to be used. If at the 50gr curve to less
pressure is in the measuring cup (LED don´t lights up orange!), than the 100gr
calibration curve must be used.
Material less than 15% water content cannot be measured.
Information: It is possible that other calibration curves (e.g., other types of wood,
defined mixtures, etc.) are calibrated by the Schaller GmbH in the humimeter
device.