Configuration – Checkline CAP-TA User Manual
Page 7

User’s Guide
CONFIGURATION
CAP-TA torque testers have several features with programmable options allow-
ing many user-specified choices. To enter the configuration menu, perform the
following:
1. Turn off the tester
2. Press and hold PEAK
3. Turn on the tester
4. Release PEAK
The version number of the internal software will be displayed for a short time.
The following secondary functions of keys are used during the configuration
process:
ENTER
Used to select a menu choice
ADVANCE
Used to step through menu choices
NOTE: Once the menu has been entered, it can only be exited if changes have
been made.
The following list shows all configuration options. Italics indicate factory set-
tings.
232 - RS-232 settings sub-menu
232d
Output
Disabled
232E Output
Enabled
300
300
baud
600
600
baud
1200
1200
baud
2400
2400
baud
4800
4800
baud
9600 9600
baud
7-1E
7 data bits, 1 stop bit, even parity
7-1o
7 data bits, 1 stop bit, odd parity
7-2E
7 data bits, 2 stop bits, even parity
7-2o
7 data bits, 2 stop bits, odd parity
7-2n
7 data bits, 2 stop bits, no parity
8-1E
8 data bits, 1 stop bit, even parity
8-1o
8 data bits, 1 stop bit, odd parity
8-1n
8 data bits, 1 stop bit, no parity
8-2n
8 data bits, 2 stop bits, no parity
Ft F
Full data (numeric + units)
Ft n
Numeric data only
bcd - Mitutoyo BCD settings sub-menu
bcdd
Output
disabled
bcdE Output
enabled
nPOL
No polarity (absolute value)
POL
Data with polarity (+ for CW, - for CCW)
5