Imada dtx2 series specifications preparation, Precautions – Checkline DTX2 User Manual

Page 3

background image

page 3

page 14

Accuracy

± 0.5% F.S. ± 1 LSD

Selectable Units

lbf-in, kgf-cm, and N-cm

Overload Capacity

200% of F.S. Overload indicator flashes beyond 110% of F.S.

Power

Rechargeable Ni-MH battery pack or Imada AD120/230 adapter

Battery Indicator

Icon flashes when battery is low

Battery Life

approx. 8 hours (recharge time approx. 10 hours)

Memory

Non-volatile, recall up to 1,000 data

Setpoints

Programmable high/low setpoints with color-coded LED
indicators and output signal

Outputs

RS-232C, Digimatic and ± 2 VDC analog output

Auto Power Off

5, 10, 30, 60 minutes or OFF (selectable)

Operating Temp.

32° to 100°F (0° to 40°C)

Imada DTX2 Series Specifications

PREPARATION

The sample table (C) must be assembled with the main unit.
1. Connect AC adapter/charger (AD120) to the tester, then plug

into 115 VAC outlet (AD230 AC adapter/charger for 230VAC).
Tester can also use inter-
nal Ni-MH batteries
without plugging into
AC power. Turn on the
tester and check that the
display is at zero and
overload indicator is
not flashing.

PRECAUTIONS

DTX2 Cap Torque Tester is a sensitive instrument. The load cell
can be damaged regardless of whether the unit is ON or OFF.
Follow these precautions to keep your DTX2 from being damaged.
1. Do not exceed unit’s capacity regardless if power is on or off.
2. Only use the Imada AC adapter/charger, other brands may cause

serious damage.

3. Accuracy may be affected if unit is exposed to high humidity,

dust or extreme shock.

4. Do not disassemble unit. Disassembly voids warranty.
5. Recommended re-calibration cycle is one (1) year.
6. When operating or moving do not lift or twist the DTX2 table

or carry by holding the knob or table.

7. Before shipping, remove the DTX2 table from the base and

pack both items together.

(A) CLAMP HANDLE

(C) SAMPLE TABLE

SHAFT WITH KEYWAY

OPENING WITH KEY

CAUTION:

Do not put any foreign
object in the opening.
Do not put excess force
or weight on sample
table (C).