DIMENSIONS

Model

Capacity (Resolution)

lbf-in

kgf-cm

N-cm

DTX2-15

15.00 (0.01 lbf-in)

20.00 (0.01 kgf-cm)

200.0 (0.1 N-cm)

DTX2-40

40.00 (0.01 lbf-in)

50.00 (0.01 kgf-cm)

500.0 (0.1 N-cm)

DTX2-85

85.0 (0.1 lbf-in)

100.0 (0.1 kgf-cm)

1000 (1 N-cm)

INTRODUCTION

Model DTX2 is a highly sophisitcated, cap torque tester which offers
programmable high/low setpoints for go/no go testing. Store up to
1,000 values into memory, which can be transmitted using Digimatic
or RS-232 formats. Use the Real Time mode to display torque tran-
sients, or the Peak mode to capture the peak torque achieved during
a test. Select measuring units between lb-in, kg-cm and N-cm.

Programmable Setpoint LED’s
When high-low setpoints are set, LED
indicates below (-NG), within (OK),
or above set point value (+NG).

CW icon
Displayed for CW measurements.

CCW icon
Displayed for CCW measurements.

Battery icon
Flashes when the NI-MH cells need
to be charged.

PEAK icon
When continuously displayed, peak
function is activated.

HOLD icon
Displays when external hold signal
is activated or SEND button is
pressed.

Torque Units
Displays selected measuring units.
(Lbf-in, kgf-cm and N-cm)

Overload Indicator
Flashes at 110% of rated capacity.

Alarm Icon

Auto Power Off icon

10

9

8

7

6

5

4

3

2

1

(A) CLAMP HANDLE

(E) AC ADAPTER/CHARGER RECEPTACLE

(ON BACK)

(B) 1" TALL HOLDING PEGS (CUSTOM

HEIGHT PEGS ARE AVAILABLE)

(C) SAMPLE TABLE

(D) COMMUNICATIONS PORT

RS-232C, Digimatic & ± 2VDC Analog

1

2

3

4

5

6

7

8

9 10