Checkline TI-44N User Manual
Page 5
![background image](/manuals/656288/5/background.png)
4.0
G
ETTING
S
TARTED
4.1 Setting Units Of Measure
Remove the Battery Cover on the back side of the gauge
by pressing it down with your thumb and sliding it in the
direction indicated by the arrow. The Units Selector Switch
is located in the Battery Compartment. Slide the switch to
either the I n c h (up) or m m (down) position as desired.
Inch is the factory default unit of measure .
4.2 Installing The Battery
Remove the Battery Cover on the back side of the gauge by pressing it down
with your thumb and sliding it in the direction indicated by the arrow. Place
one (1) AA-size battery (1.5 Volt) into the battery compartment in the
orientation shown on the sketch. Replace the Battery Cover.
AA Battery
mm
Inch
+
–
Correct Battery Orientation
Battery
Compartment
TI-44N Operating Instruction Guide
5
Description of Problem
Possible Cause
Action To Be Taken
Fluxuating readings
Defective probe
Return to manufacturer
Fluxuating readings while
Aluminum measurement
Switch to aluminum mode
measuring aluminum
is being taken in standand
mode
No reading or wildly
Measurment is being taken Switch to R to B1 mode
fluxuating reading while
in S to B1 mode
measuring resin
Flashing coupling signal
Probe or meter failure or
Return to manufacturer for
and/or reading is displayed
deterioratioin
service
while the measurement is
in standby
Sonic velocity reading
• Effect of external noise
• Change use location
cannot be modified (fixed
to 5930 m/s)
• Meter failure
• Return to manufacturer for
service
Measurement of Test
• Accoustic Velocity not
• Change velocity to 5930
Plate does not show
set to 5930 (steel)
5930 µs.
0.197”
(5.00mm)
(Refer to section 5.0)
TI-44N Operating Instruction Guide