4.1 Setting Units Of Measure

Remove the Battery Cover on the back side of the gauge
by pressing it down with your thumb and sliding it in the
direction indicated by the arrow. The Units Selector Switch
is located in the Battery Compartment. Slide the switch to
either the I n c h (up) or m m (down) position as desired.
Inch is the factory default unit of measure .

4.2 Installing The Battery

Remove the Battery Cover on the back side of the gauge by pressing it down
with your thumb and sliding it in the direction indicated by the arrow. Place
one (1) AA-size battery (1.5 Volt) into the battery compartment in the
orientation shown on the sketch. Replace the Battery Cover.

AA Battery

mm

Inch

+

Correct Battery Orientation

Battery
Compartment

TI-44N Operating Instruction Guide

5

Description of Problem

Possible Cause

Action To Be Taken

Fluxuating readings

Defective probe

Return to manufacturer

Fluxuating readings while

Aluminum measurement

Switch to aluminum mode

measuring aluminum

is being taken in standand
mode

No reading or wildly

Measurment is being taken Switch to R to B1 mode

fluxuating reading while

in S to B1 mode

measuring resin

Flashing coupling signal

Probe or meter failure or

Return to manufacturer for

and/or reading is displayed

deterioratioin

service

while the measurement is
in standby

Sonic velocity reading

• Effect of external noise

• Change use location

cannot be modified (fixed
to 5930 m/s)

• Meter failure

• Return to manufacturer for

service

Measurement of Test

• Accoustic Velocity not

• Change velocity to 5930

Plate does not show

set to 5930 (steel)

5930 µs.

0.197”

(5.00mm)

(Refer to section 5.0)

TI-44N Operating Instruction Guide