Checkline TI-CMXDL User Manual
Page 24
-24-
Material
sound velocity
sound velocity
in/us
m/s
Aluminum
0.2510
6375
Beryllium
0.5080
12903
Brass
0.1730
4394
Bronze
0.1390
3531
Cadmium
0.1090
2769
Columbium
0.1940
4928
Copper
0.1830
4648
Glass (plate)
0.2270
5766
Glycerine
0.0760
1930
Gold
0.1280
3251
Inconel
0.2290
5817
Iron
0.2320
5893
Cast Iron
0.1800
(approx)
4572
Lead
0.0850
2159
Magnesium
0.2300
5842
Mercury
0.0570
1448
Molybdenum
0.2460
6248
Monel
0.2110
5359
Nickel
0.2220
5639
Nylon
0.1060
(approx)
2692
Platinum
0.1560
3962
Plexiglas
0.1060
2692
Polystyrene
0.0920
2337
PVC
0.0940
2388
Quartz glass
0.2260
5740
Rubber vulcanized
0.0910
2311
Silver
0.1420
3607
Steel (1020)
0.2320
5893
Steel (4340)
0.2330
5918
Steel Stainless"
0.2230
5664
Teflon
0.0540
1372
Tin
0.1310
3327
Titanium
0.2400
6096
Tungsten
0.2040
5182
Uranium
0.1330
3378
Water
0.0580
1473
Zinc
0.1660
4216
Zirconium
0.1830
4648
Acoustic Velocity Table
10.0 ADDITIONAL FEATURES OF THE TI-CMXDL
10.1 High Speed Scan
The High Speed Scan feature of the TI-CMXDL increases the overall repetition rate to a
maximum of 140Hz with a high speed screen refresh rate of 25 times a second. This
feature enables a user to make scanned passes over an arbitrary length of the test materi
while still maintaining a reasonable representation of thickness over the scanned area or
region. This feature can be used in conjunction with High and Low alarm limits features
to dynamically keep track of both values. The feature is typically used to provide a bette
representation of the area scanned, by taking more readings at a faster rate repetition rat
The procedure to use the SCAN MODE feature is outlined below:
Using the Scan Mode
1. Press the MENU key once to activate the menu items tab.
Press the MENU key multiple times to tab right and the ESC
key multiple times to tab left until the UTIL menu is highlighted
and displaying the submenu items.
2. Use the UP and DOWN arrow keys to scroll through the sub
menu items until SCAN MODE is highlighted.
3. Use the LEFT and RIGHT arrow keys to toggle the
SCAN MODE on/off.
4. Press the MEAS key to return to the measurement screen.
10.2 Alarm Mode
The Alarm Mode feature of the TI-CMXDL provides the user with a method of setting t
erances, or limits, for a particular application requirement. This feature may be used for
variety of applications to verify the material is within the manufacturer specifications.
There are two limits, or alarm values, that can be setup in the TI-CMXDL – ALARM LO
and ALARM HIGH limits. However, the user may choose to activate and utilize only on
of the limit values, depending on their specific application requirements. The procedures
use the ALARM MODE feature are outlined below:
Toggle Alarm (on/off)
1. Press the MENU key once to activate the menu items tab.
Press the MENU key multiple times to tab right, and the
ESC key multiple times to tab left, until the UTIL menu is
highlighted and displaying the submenu items.
2. Use the UP and DOWN arrow keys to scroll through the
sub menu items until ALARM is highlighted.
3. Use the LEFT and RIGHT arrow keys to toggle the
ALARM on/off/audible.
4. Continue on to the next section “Setting the Alarm Low
Limi.t."
-45-