Cannon Instrument miniAV-LT User Manual

Page 57

background image

53

CANNON

®

miniAV-LT Automatic Viscometer with VISCPRO

®

Instruction & Operation Manual

Version 1.a— May, 2009; CANNON

®

Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

Enter the maximum acceptable difference in measured results when using
multiple determinations (the default percent difference value is 0.35
percent and the default absolute difference value is 0.2 seconds).

Use greater

If both tolerance options are selected, the VISCPRO software will apply
both criteria unless the Use greater option is checked. If Use greater is
checked, then the software will apply the less exacting tolerance value
when calculating/evaluating test results.

The Tray Settings: Test window

Determinations

Select the determination option by clicking on the radio button corre-
sponding to the appropriate determination method (Always perform
maximum determinations, Use successive determinations only, or
Use any determinations).

Selecting sample run required and maximum determination settings

Select the number of Required determinations by clicking on the
spin controls (values from 1 to 4 are acceptable). In the same manner,
select the Maximum number of determinations which can be run for the
tube samples (values from 1 to 20 are acceptable).

NOTES

VISCPRO determines the sample flow time by averaging the values of
the sample runs which correspond to the chosen criteria.