miniAV-LT calibration equations

Viscosity calculation equation In the miniAV-LT, viscosity is calculated by using the viscosity calcula-

tion equations as described in ASTM Methods D 445 and D 446. Because
samples running in the miniAV-LT can have short flow times, the kinetic
energy correction term (E / t

2

) is used as prescribed in the aforementioned

ASTM methods. The VISCPRO

®

software for Windows

®

performs all

necessary calculations to derive constants C and E. These constants are
used by the software when determining values for kinematic viscosity
using sample drop times measured by the miniAV-LT.

Kinematic Viscosity equation: v = C t - E / t

2

Where: v

=

kinematic viscosity (cSt)

C

=

calibration constant (cSt/s)

E

=

kinetic energy correction (cSt·s

2

)

t

=

flow time (seconds)

NOTE

When calibrating the miniAV-LT, the operator should analyze the viscos-
ity standard using double determinations to obtain two flow times for
each standard. The average of the two flow times is used as input data
for the calibration program.