6 suction module, removal – BUCHI KjelDigester K-449 User Manual
Page 54
7 – 54
7 Maintenance
KjelDigester K-446/K-449
Operation Manual
7.6 Suction module, removal
1. Disconnect the Scrubber hose c from the fume collection tube and store it the
bracket left of the device.
2. Hold the suction module a on both handles b and lift it up and to the rear.
NOTE
The magnets on rack support will hold the suction module back until the force of the mag-
netic field is overcome by lifting up.
CAUTION
Risk of chemical burns from acid!
Minor or moderate cuts can result.
1. Wear the personal protective equipment.
3. Store and carry the suction module upside down to prevent acid drops pour out of
the suction module.