beautypg.com

4 set parameters for an automatic digestion – BUCHI KjelDigester K-449 User Manual

Page 43

background image

6 – 43

Operation Manual

KjelDigester K-446/K-449

6 Operation

6.8.4 Set parameters for an automatic digestion

1. Press Method [METH].

2. Select a method number between 1 and 9.

6.8.6 Select a method, (6–44).

3. Press c [STEP].

∙The text [STARTTIME] on the display appears.

4. Set a delayed start time if required or press c [STEP] to continue.

6.8.7 Set up a delayed start time, (6–44).

5. Press a [TEMP] and adjust the temperature by pressing d [+/-].
6. Press e [ENTER] to confirm the new value.

7. Press b [TIME] and adjust the time by pressing d [+/-].

8. Press e [ENTER] to confirm the new value.
9. Press c [STEP].

10. Repeat the steps 5 – 9.

NOTE
If not all of the 4 steps are required, set the time to zero on the remaining steps.

11. Press a [TIME] and adjust the cooling time to keep the Scrubber running after the

digestion is finished by pressing c [+/-].
NOTE
Set up a time amount big enough that no more fumes are evading from the sample tubes
when the Scrubber stops. BUCHI recommends a starting value of 30 minutes.

12. Press d [ENTER] to confirm the new value.

13. Press b [STEP].

This manual is related to the following products: