BUCHI KjelDigester K-449 User Manual
Page 28
5 – 28
5 Installation
KjelDigester K-446/K-449
Operation Manual
b
Tasks
WARNING
Risk of acid or toxic fume inhalation.
Serious injuries or death can result.
1. Always run the Scrubber or water jet pump during the digestion process.
2. Operate the Scrubber and KjelDigester only in a fume hood.
3. Close the fume hood door as far as possible.
NOTICE
Risk of wrong power input on the Scrubber.
Wrong input voltage of the Scrubber can damage electronic components.
1. Do only connect a Scrubber with 230 V input to the KjelDigester. Output volt-
age of the KjelDigester is 240 V only.
1. Connect the Scrubber connection cable on the rear of the KjelDigester.
2. Connect the hose c with the ball joint hose connector to the fume collection tube a.
3. Secure the connection with the catchpot clip b.
4. Shorten the other end of the hose c to keep the hose as short and straight as pos-
sible.
5. Connect the hose c to a Scrubber or water jet pump.
c
Follow-on tasks
1. Depending on the configuration continue with following step:
∙Install the condensate bottle (5–33).