beautypg.com

5 rechargeable battery (k-449 only), replacement – BUCHI KjelDigester K-449 User Manual

Page 52

background image

7 – 52

7 Maintenance

KjelDigester K-446/K-449

Operation Manual

7. If the fuse brakes repeatedly contact the customer support.

8.1 Customer service, (8–55).

7.5 Rechargeable battery (K-449 only), replacement

a

Required tools:

• TORX screw driver T10

b

Preliminary tasks

1. Remove the drip tray and store it on the drip tray holder on the left side of the device.

2. Remove the suction module (7–54).

c

Tasks

1. Make sure the lift is in the cooling position.

WARNING

Risk of electric shock with connected power supply cable!
Serious injuries or death can result.
1. Switch off the device.
2. Disconnect the power supply cable from the device.

2. Switch off the device.

3. Disconnect the power supply cable from the device.
4. Remove the 4 TORX screws a of the lift cover.

Required tools
TORX screw driver T10

5. Slide out the lift cover b.

This manual is related to the following products: