1 electrical system, 2 environmental conditions, 1 electrical system 3.2.2 environmental conditions – BUCHI KjelDigester K-449 User Manual
Page 18
3 – 18
3 Technical data
KjelDigester K-446/K-449
Operation Manual
Description
K-446
K-449
Display
2 row LCD display
2 row LCD display
Resolution temperature display
1 °C
1 °C
Temperature range
30 – 450 °C
30 – 450 °C
Average temperature stability
±1 °C
±1 °C
Absolute temperature accuracy
±7 °C at 420 °C
±7 °C at 420 °C
3.2.1 Electrical system
Description
K-446
K-449
Input voltage
220 – 240 V ±10 %
220 – 240 V ±10 %
Frequency
50 – 60 Hz
50 – 60 Hz
Power consumption max.
2300 W
2300 W
Output current, Scrubber
0.7 A
0.7 A
Fuse
2× T10 A/L250 V
2× T10 A/L250 V
3.2.2 Environmental conditions
Description
K-446
K-449
Temperature
5 – 40 °C
5 – 40 °C
Altitude above sea level max.
2000 m
2000 m
Humidity
Maximum relative humidity 80 % for temperatures
up to 31 °C, decreasing linearly to 50 % relative
humidity at 40 °C
Use
For indoor use only