beautypg.com

7 condensate bottle, installation – BUCHI KjelDigester K-449 User Manual

Page 33

background image

5 – 33

Operation Manual

KjelDigester K-446/K-449

5 Installation

5.7 Condensate bottle, installation

CAUTION

Risk of acid or peroxide fume inhalation!
Minor or moderate chemical burns of the airways.
1. Keep the suction hoses between the KjelDigester and Scrubber as straight

as possible to prevent liquid blocking the hoses.

2. Make sure the condensate bottle is placed lower than the suction module.
3. Operate and maintain the Scrubber according to its operation manual.
4. Make sure the hoses are not damaged.

1. Connect the hose c with the hose connector to the fume collection tube a.

2. Secure the connection with the catchpot clip b.

3. Connect the hose c coming from the KjelDigester 

B

to the condensate bottle b.

∙025100 Condensate bottle

4. Connect the hose a from the Scrubber 

A

to the condensate bottle.

5. Close the 2 remaining connections of the condensate bottle with 2 plugs.