Appendix g - communications, Appendix g, Communications – Brookfield DV-I Viscometer User Manual
Page 40
Brookfield Engineering Laboratories, Inc.
Page 40
Manual No. M/92-021-O604
Appendix G - Communications
2
3
4
5
6
7
8
9
1
No Connection
No Connection
Analog Ground (%Torque & Temperature)
Analog % Torque (
see Note 1
)
DV-I+ with Temperature Option Analog Outputs
Connected
internally. Do
not use.
No Connection
Analog Temperature - optional (
see Note 2
)
{
NOTES: 1. This is
a 0-1 volt d.c. output where 0 volts corresponds to 0% torque and 1
volt corresponds to 100% torque with a resolutuion of 1 millivolt (0.1%).
2. This is a 0-4 volt d.c. output where 0 volts corresponds to -100
°
C and 4 volts
correpsonds to +300
°
C with a resolution of 1 millivolt (0.1
°
C).
Figure G-1
Analog Output:
The analog outputs for % torque is accessed from the 9-pin connector located on the rear panel of
the DV-I+. The pin connections are shown in Figure G-1.
The output cable (Part No. DVP-96Y) connections are:
Red Wire:
(Not used with DV-I
+
)
Black Wire:
(Not used with DV-I
+
)
White Wire:
% Torque Output
Green Wire:
% Torque Ground
Please contact Brookfield Engineering Laboratories or your local dealer/distributor for purchase